Epic Test Code LAB82 Glucose, Random
Performing Location(s)
BMC,CRH,LFH,HRK,OCH,FTT,FOX,FCC
Specimen Type
Serum or Plasma
Preferred Container
Gold or Red top tube
Alternate Container
Green or Gray top tube
Blood Tube Draw Volume
Min 50% draw volume
Minimum Volume to Submit for Testing
0.5 mL Serum or Plasma
- Multiple test can be performed from this volume. For question please contact the lab at 547-3975.
Storage Requirements
Sample tubes should be centrifuged within 2 hour of collection followed by transfer of the serum (red) or plasma (green top) to a labeled plastic, aliquot tube.
Gold tops do not need to be aliquoted
See Causes of Rejection for temperature requirements.
Gray top tubes do not require centrifugation for transport up to 48 hrs.
Transportation Needs
Deliver specimen to laboratory within 1 hour of collection if unspun. If not possible see "storage Requirements".
Causes for Rejection
Mislabeled or unlabeled specimen
Less than 50% draw for Vacutainer tubes
Serum stored on clot more than 4 hours
>3 days refrigerated
> 8 hours room temperature
> 5 days Frozen
NaF >3 days Refrigerated
Hemolysis > 500 mg/dL HgB
Lipemia > 700 mg/dL Trig
Bilirubin > 40 mg/dL Bili
Limitations
Hemoglobin of 500 mg/dL [0.31 mmol/L (monomer)]d falsely depresses GLU results by 20 mg/dL [1.12 mmol/L] at 131 mg/dL [7.27 mmol/L] of glucose.
Bilirubin of 20 mg/dL [342 µmol/L] falsely depresses GLU results by 12 mg/dL [0.7 µmol/L] at 124 mg/dL [6.9 µmol/L of glucose.
Lipemic (triglyceride 600 mg/dL [6.86 mmol/L]) samples falsely elevate GLU results by 10 mg/dL [0.56 mmol/L] at 129 mg/dL [7.16 mmol/L] of glucose.
Capillary collected blood may differ from venous due to contamination from hemolysis or interstitial fluid.
Reference Values
Random Glucose Reference Ranges |
|||
Age | Value | High Critical Value | Low Critical Value |
<1 week | 45-100mg/dL | > 450 | < 45 |
≥1 week | 70-139 mg/dL | > 450 | <40 |
The diagnosis of diabetes can be confirmed by: |
|
** In the absence of unequivocal hyperglycemia, these criteria should be confirmed by repeat testing from a laboratory on a different day. |
Available STAT
Yes
Days of Analysis
All
Methodology
Beckman AU/Dx Series
CPT Code
82947
Last Updated
27-NOV-24 MS