Epic Test Code LAB2448 Cytomegalovirus (CMV) Molecular Detection, PCR, Varies
Additional Codes
MML:CMVPV
Ordering Guidance
For plasma specimens order CMVQN / Cytomegalovirus (CMV) DNA Detection and Quantification by Real-Time PCR, Plasma.
Necessary Information
Specimen source is required.
Specimen Required
Submit only 1 of the following specimens:
Specimen Type: Body fluid
Sources: Spinal, pleural, peritoneal, ascites, pericardial, amniotic, or ocular
Supplies: Sarstedt Aliquot Tube, 5mL (T914)
Container/Tube:
Preferred: Sterile, screwcap, 5-mL aliquot tube
Acceptable: Sterile container
Specimen Volume: 0.5 mL
Collection Instructions: Do not centrifuge.
Specimen Type: Respiratory fluid
Sources: Bronchial washing, bronchoalveolar lavage, nasopharyngeal aspirate or washing, sputum, or tracheal aspirate
Supplies: Sarstedt Aliquot Tube, 5mL
Container/Tube:
Preferred: Sterile, screwcap, 5-mL aliquot tube
Acceptable: Sterile container
Specimen Volume: 1.5 mL
Collection Instructions: Do not centrifuge.
Specimen Type: Genital swab
Sources: Cervix, vagina, urethra, anal/rectal, or other genital sources
Supplies:
-Culturette (BBL Culture Swab) (T092)
-M4-RT (T605)
Container/Tube: Multimicrobe media (M4-RT) and ESwabs
Collection Instructions: Place swab back into multimicrobe media (M4-RT, M4, or M5)
Specimen Type: Swab
Sources: Dermal, eye, nasal, saliva, or throat
Supplies:
-Culturette (BBL Culture Swab) (T092)
-M4-RT (T605)
Container/Tube: Multimicrobe media (M4-RT) and ESwabs
Collection Instructions: Place swab back into multimicrobe media (M4-RT, M4, or M5)
Specimen Type: Tissue
Sources: Brain, colon, kidney, liver, lung, etc.
Supplies: M4-RT (T605)
Container/Tube: Sterile container containing 1 mL to 2 mL of sterile saline or multimicrobe medium (M4-RT, M4, or M5)
Specimen Volume: Entire collection
Collection Instructions: Submit only fresh tissue in multimicrobe media (M4-RT) or a sterile container with 1 to 2 mL sterile saline
Specimen Type: Urine
Container/Tube: Sterile container
Specimen Volume: 1 mL
Collection Instructions: Collect a random urine specimen.
Specimen Type: Bone marrow
Container/Tube: Lavender top (EDTA)
Specimen Volume: 0.5 mL
Collection Instructions: Send bone marrow in original tube. Do not aliquot.
ACCEPTABLE SWABS: Swabs in media, such as M4-RT, M4 or M5, Bartels FlexTrans, Jiangsu VTM, BBL CultureSwab, E-Swabs with media –
(dry/flocked E-Swabs are NOT acceptable).
Useful For
Rapid qualitative detection of cytomegalovirus (CMV) DNA
This test is not intended for the monitoring of CMV disease progression.
Method Name
Real-Time Polymerase Chain Reaction (PCR)/DNA Probe Hybridization
Reporting Name
Cytomegalovirus, PCRSpecimen Type
VariesSpecimen Minimum Volume
Ocular Fluid, Spinal Fluid, or Urine: 0.3 mL
Body Fluid (pleural, peritoneal, ascites, pericardial): See Specimen Required
Respiratory Specimens: 1 mL
Tissue: 2 × 2-mm biopsy
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Varies | Refrigerated (preferred) | 7 days | |
Frozen | 7 days |
Reject Due To
Calcium alginate-tipped swab Wood swab Transport swab containing gel Feces Paraffin blocks Breast milk Heat inactivated specimens |
Reject |
Clinical Information
Infection with cytomegalovirus (CMV) is a significant cause of morbidity and mortality in transplant recipients and other immunocompromised hosts. Specific neurologic syndromes associated with CMV infection include subacute radiculomyelopathy, peripheral neuropathy, and encephalitis.
CMV-associated central nervous system (CNS) disease occurs most commonly in immunocompromised patients. Histologic evidence of CMV infections in autopsy brain tissue was identified in 20% to 40% of AIDS patients. In 2 separate studies, CMV (DNA) was the most common herpesvirus (29/181, 16/49) detected from the cerebrospinal fluid of patients with AIDS.
CNS infections with CMV can also occur in immunocompetent patients. CMV is a leading cause of congenital viral infections worldwide, and laboratory testing by real-time polymerase chain reaction is useful in the diagnosis of neonatal CMV disease.
Reference Values
Negative
Reference values apply to all ages.
Interpretation
Detection of cytomegalovirus (CMV) DNA in a specimen supports the clinical diagnosis of infection due to this virus.
Studies indicate that CMV DNA is not detected by polymerase chain reaction in cerebrospinal fluid from patients without central nervous system disease caused by this virus.
Cautions
A negative result does not eliminate the possibility of cytomegalovirus (CMV) infection.
This assay is only to be used for patients with a clinical history and symptoms consistent with CMV infection and must be interpreted in the context of the clinical picture.
Supportive Data
The following validation data support the use of this assay for clinical testing.
Accuracy:
A total of 200 prospective clinical samples (respiratory [n=72], urine [n=67], spinal fluid [n=25], fresh tissue [n=18], amniotic fluid [n=10], and bone marrow [n=8]) were submitted to our reference laboratory for cytomegalovirus (CMV) real-time polymerase chain reaction (PCR) (Roche analyte specific reagents, Roche Diagnostics). Respiratory samples included bronchoalveolar lavage (BAL) fluid (n=25), bronchial washing (n=40), nasal swab (n=4), tracheal secretions (n=2), and throat swab (n=1). Each sample was tested by 6 real-time PCR assays, and the results were compared to consensus reference standard (4 of 6 results being in agreement). The performance of the US9 CMV real-time PCR (laboratory-developed test) is summarized in the Table:
Table. Performance of the US9 CMV real-time PCR assay following testing of prospective clinical samples (n=200)
US9 CMV PCR |
Consensus result |
||||
Positive |
Negative |
Kappa |
Sensitivity |
Specificity |
|
Positive |
45 |
0 |
0.99 |
97.8 (87.6-99.9) |
100 (97.1-100) |
Negative |
1 |
154 |
|
|
|
Analytical Sensitivity/Limit of Detection:
To evaluate the analytical sensitivity, whole virus control (Acrometrix, Life Technologies) at a starting concentration of 500,000 copies/mL was used to generate a dilution panel. In brief, samples were diluted 1:2 in tris-EDTA buffer to a final concentration of 8 copies/mL. Each member of the dilution panel was then tested in triplicate, with the limit of detection (LOD) being defined as the highest dilution at which all replicates tested positive. The LOD was determined to be 122 copies/mL.(1)
Analytical Specificity:
No PCR signal was obtained from extracts of 44 bacterial and viral isolates including Epstein-Barr virus, herpes simplex virus, varicella-zoster virus, human herpes virus (HHV) 6, HHV7, HHV8, and parvovirus.
Day(s) Performed
Monday through Sunday
Report Available
Same day/1 to 4 daysSpecimen Retention Time
7 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterCPT Code Information
87496
NY State Approved
YesForms
If not ordering electronically, complete, print, and send a Microbiology Test Request (T244) with the specimen.