Sign in →

Epic Test Code LAB99 Lipase: Serum

Performing Location(s)

BMC,FOX,FTT,CRH,LFH,OCH,FCC,HRK

Specimen Type

Serum or Plasma

Preferred Container

Gold or Red top tube

Alternate Container

Green top tube

Minimum Volume to Submit for Testing

0.5 mL Serum or Plasma

  • Multiple test can be performed from this volume. For question please contact the lab at 547-3975.  

Transportation Needs

 Deliver specimen to laboratory within 1 hour of collection if unspun. If not possible see "storage Requirements". 

Storage Requirements

Sample tubes should be centrifuged within 2 hours of collection 

Gold tops are stable in original tubes for 7 days DO NOT ALIQUOT

Aliquot Red tops

See Causes of Rejection for temperature requirements.

Causes for Rejection

Mislabeled or unlabeled specimen
Less than 50% draw for Vacutainer tubes
 >7 days refrigerated 
> 24 hours room temperature
> 1 year Frozen

Hemolysis > 500 mg/dL HgB
Lipemia > 500 mg/dL Trig
Bilirubin > 12 mg/dL Bili

EDTA, NaF, Citrate, Oxalate

Limitations

Bacterial contamination of the specimen may cause increased lipase values

Hemoglobin at a test concentration of 500 mg/dL [0.31 mmol/L] increases the lipase activity by 30 U/L at a lipase level of 143 U/L. Hemoglobin at a test concentration of 500 mg/dL [0.31 mmol/L] has less than 10% interference at lipase levels greater than 300 U/L.

EDTA, potassium oxalate, sodium fluoride and citrate have been shown to inhibit lipase results and should not be used.

Reference Values

Population Units Reference Range
All populations U/L 11-82

 

Days of Analysis

All

Available STAT

Yes

Additional Information

Contraindications:

Lipase is increased in approximately 50% of patients with chronic renal failure. Serum lipase increases with hemodialysis. Therefore, predialysis blood samples are recommended for lipase movement.

Methodology

Beckman AU/Dx Series

CPT Code

83690

Last Updated

26-OCT-24 MS